Installation and Operation Manual

Thermo Scientific Revco

®

 

Ultima II

Cryogenic Storage Freezers

 


background image

©

 2007 Thermo Fisher Scientific. All rights reserved.

“Suva

®

” is a registered trademark of DuPont.

“Mylar

®

” is a registered trademark of DuPont.

“Armaflex

®

” is a registered trademark of Armacell International GmBH.

All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.

2007 Thermo Fisher Scientific. All rights reserved.

“Suva

®

” is a registered trademark of DuPont.

All other trademarks are the property of Thermo Fisher Scientific Inc. and 

its subsidiaries.

 


background image

Thermo Scientific Revco Cryogenic Storage Freezers

Installation and Operation

i

Table of Contents

1 About Cryogenic Models

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  1

2 Safety Precautions

 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  1

3 General Recommendations

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  1

3.1 Initial Loading

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

4 Pre-Installation

 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  1

4.1 Unpacking

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

5 Installation

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  2

5.1 Location

 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2

5.2 Wiring

 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2

5.3 Leveling

 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2

6 Operation (Revco Ultima II Freezers)

 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  3

6.1 Control Panel Features

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

6.2 Start Up

 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

6.2.1

Turning the Power On

 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

6.2.2

Setting the Cabinet Temperature

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

6.2.3

Setting the Alarms

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

6.2.4

Alarm Reset & Status Lights

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

6.2.5

Alarm Test

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

6.3 Backup System (Optional)

 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

6.3.1

LN

2

 Precautions

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

6.3.2

Installation

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

6.3.3

Operation

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

6.3.4

System Status Lights

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

7 Maintenance and Troubleshooting

 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  7

7.1 Condenser Maintenance

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

7.1.1

Cleaning the Condenser

 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

7.1.2

Cleaning the Condenser Filter

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

7.2 Gasket Maintenance

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

7.3 Defrost Procedures

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

7.4 Alarm Battery Maintenance

 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

7.5 Frame Heater Switch

 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

7.6 Troubleshooting Procedures

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

 


background image

Thermo Scientific Revco Cryogenic Storage Freezers

Installation and Operation

ii

8 Chart Recorders

 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  8

8.1 Set Up and Operation

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

8.2 Power Supply

 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

8.3 Changing Chart Paper

 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

8.4 Calibration Adjustment

 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

 


background image

Thermo Scientific Revco Cryogenic Storage Freezers

Installation and Operation

1

1 About Cryogenic Models

The freezers described in this manual are 

cryogenic

 models. 

These models are designed for optimum performance at -140ºC 
or -150ºC, and it is strongly recommended that you use these 

models at these operating temperatures.

Be sure to read the maintenance instructions in Section 7.

2 Safety Precautions

In this manual and on labels attached to this product, the words 
WARNING and CAUTION mean the following:

WARNING: a potentially hazardous situation which, if not 
avoided, could result in serious injury or death.

CAUTION: a potentially hazardous situation which, if not 
avoided, may result in minor or moderate injury or damage to 
the equipment.

Before installing, using or maintaining this product, please be sure 
to read this manual and product warning labels carefully. Failure to 
follow these instructions may cause this product to malfunction, 
which could result in injury or damage.

Below are important safety precautions that apply to this product:

Use this product only in the way described in the product 
literature and in this manual. Before using it, verify that this 
product is suitable for its intended use. 

Do not modify system components, especially the controller. 
Use OEM exact replacement equipment or parts. Before use, 
confirm that the product has not been altered in any way.

Your unit must be properly grounded in conformity with 
national and local electrical codes. Never connect the unit to 
overloaded power sources.

Disconnect the unit from all power sources before cleaning, 
troubleshooting, or performing other maintenance on the 
product or its controls.

3 General Recommendations

The refrigeration system is designed to maintain cryogenic 
temperatures with safety 

only

 when the freezer is used for 

storage. These models may be used in an ambient environment 
up to +25ºC (77ºF). 

WARNING!

This unit is not a “rapid-freeze” device. 

Freezing large quantities of liquid, or high-water 
content items, will temporarily increase the chamber 
temperature and will cause the compressors to 
operate for a prolonged time period. 

Avoid opening the door for extended time periods since chamber 

temperature air will escape rapidly. Room air, which is higher in 
humidity, replacing chamber air may cause frost to develop in the 
chamber more rapidly.

3.1 Initial Loading

After reading and completing the Safety Considerations, 
Pre-Installation, Installation, and Operation sections of this 
manual, turn the Key Switch to the POWER ON position and 
adjust the temperature setpoint if necessary.

The setpoint should be no warmer than -50ºC. 

Allow the freezer 

to operate at the desired temperature for a minimum of 12 
hours before loading.

CAUTION!

IFailure to follow these procedures or 

overloading the unit may cause undue stress on the 
compressors or jeopardize user product safety.

4 Pre-Installation

4.1 Unpacking

At delivery, examine the exterior for physical damage while the 

carrier’s representative is present. If exterior damage is 
present, carefully unpack and inspect the unit and all 
accessories for damage.

If there is no exterior damage, unpack and inspect the 
equipment within five days of delivery. If you find any damage, 
keep the packing materials and immediately report the damage 

to the carrier. 

Do not return goods without written 

authorization

. When submitting a claim for shipping damage, 

request that the carrier inspect the shipping container and 

equipment.

CAUTION!

Do not discard the sublids from 

chest-style units.

 The sublids are necessary to 

maintain correct temperature, moisture control, and 
economy of operation. 

 


background image

Thermo Scientific Revco Cryogenic Storage Freezers

Installation and Operation

2

5 Installation

Do not exceed the electrical and temperature ratings printed on 
the dataplate located on the lower left side of the unit.

CAUTION!

Improper operation of the equipment 

could result in dangerous conditions. To preclude 
hazard and minimize risk, follow all instructions and 
operate within the design limits noted on the dataplate.

5.1 Location

Install the unit in a level, well-ventilated area free from vibration, 

with a minimum of six inches of space on the sides and 12 inches 
behind the rear. Refer to Section 5.3 for further instructions on 
leveling cabinets. Allow enough clearance so that the lid can 

swing open at least 90 degrees.

Do not position the equipment in direct sunlight or near heating 
diffusers, radiators, or other sources of heat. The ambient 

temperature range at the location must be 59 to 77ºF 
(15 to 25ºC). 

CAUTION!

To allow for proper air flow, a minimum 

of 12 inches of clearance space is required behind the 
freezer. 

CAUTION!

FLAMMABLE REFRIGERANT! 

Cryogenic models with voltage M, R, V and W use 
small quantities of the refrigerant R600a, which can 
form a combustible mixture when combined with 
oxygen. In order to conform with safety regulations, it 
is important to locate your freezer in a large enough 
space to minimize danger of combustion even in 
worst-case conditions (no ventilation, no air exchange, 
a steady refrigerant leak). Under these conditions the 
minimum free volume requirements would be at least 
86 m

3

 or 3006 ft

3

.

5.2 Wiring

CAUTION!

Connect the equipment to the correct 

power source. Incorrect voltage can result in severe 
damage to the equipment.

CAUTION!

For personal safety and trouble-free 

operation, this unit must be properly grounded before 
it is used. Failure to ground the equipment may cause 
personal injury or damage to the equipment. Always 
conform to the National Electrical Code and local 
codes. Do not connect the unit to overloaded power 
lines.

Your freezer is equipped with the plug style shown in Figure 1, 

unless it is a 50Hz or 3-phase model. The standard NEMA plug 
meets UL requirements.

Figure 1.  Plug Style for 60Hz Single-phase Models

Always connect the freezer to a dedicated (separate) circuit. Each 
freezer is equipped with a service cord and plug designed to 
connect it to a power outlet which delivers the correct voltage. 

Supply voltage must be within +10%, -5% of the freezer rated 
voltage.

3-phase or 50Hz models require hard wiring to a disconnect 
panel.

Note:

RevcoUltima freezers are equipped with a standard 
built-in Voltage Safeguard circuit to detect and adjust 
low line voltage.

CAUTION!

Never cut the grounding prong from the 

service cord plug. If the prong is removed, the 
warranty is invalidated.

5.3 Leveling

The unit must be level both front to back and side to side. 

To level chest models, leave the casters in place and shim the low 
wheel(s) with strips of sheet metal cut at least 1/2 in. wider than 

the caster.

30A-250V

NEMA 6-30P

 


background image

Thermo Scientific Revco Cryogenic Storage Freezers

Installation and Operation

3

6 Operation 

6.1 Control Panel Features

Figure 2.  Revco Ultima II Series Control Panel with 

Optional Backup System

Before the initial start up, take some time to become familiar 

with the controls on your freezer. Figure 2 illustrates the Revco 
Ultima II Cryogenic Freezer control panel.

Note:

Figure 2 shows all possible functions. These functions 

vary depending on the options. You may not have all of 
the functions on your equipment. The CO

2

 or LN

2

 

Backup System lights (20, 21, 22) represent status 
indicators for a built-in CO

2

 or LN

2

 

backup system, 

which is optional

.

1. Three position keyed Power On Switch.
2. Power On Indicator. This indicator lights when power is 

connected to the freezer.

3. Temperature Failure Indicator. This indicator lights when the 

freezer temperature deviates either above or below the alarm 
temperature settings.

4. Power Failure Indicator. This indicator lights when there is a 

power failure to the freezer (refer to Section 7 on page 7).

5. Life-Guard Alert Indicator. This indicator lights when the 

first stage compressor has gone into a cautionary state of 
operation to help prevent failure.

6. Check Fuse Light. This indicator lights when the surge 

protector on the main electronic panel is blown by a power 
surge (refer to Section 7 on page 7).

7. Alarm Test Pad. Press this pad to start an alarm test (refer to 

Section 5.2.5 on page 6).

8. Alarm Reset Pad. Press this pad to reset

 slowly

 blinking 

indicators. Slowly blinking indicators denote that the 

condition 

has

 occurred but is now within the given operating 

parameters. This pad is also depressed to temporarily silence 
the audible alarm (refer to Section 5.2.4 on page 5).

9. Cold Alarm Setpoint. Press this pad to set the cold alarm 

(refer to Section 6.2.3 on page 4).

10. Warm Alarm Setpoint. Press this pad to set the warm alarm 

(refer to Section 6.2.3 on page 4). 

11. Control Setpoint. Press this pad to set the cabinet temperature 

(refer to Section 6.2.2 on page 4).

12. Cabinet Temperature Indicator. This indicator lights when the 

temperature display window is showing the cabinet 
temperature (refer to Section 6.2.1 on page 4).

13. Digital Temperature Display Window. This window displays 

chamber temperature, alarm values, etc. depending on the 
operating status of the freezer and the procedure being 
performed.

14. Decrement Pad (

). Use this pad to decrease temperature 

values.

15. Increment Pad (

). Use this pad to increase temperature 

values.

16. Extreme Ambient Alert. This indicator lights when the 

ambient temperature has exceeded the upper limit of the 
recommended operating range. If the Extreme Ambient Alert 

is activated, the ambient environment needs to be improved 
for proper function of the freezer.

17. Clean Filter Indicator. This indicator lights when the air 

cooled condenser filter is dirty (refer to Section 7 on page 7).

18. Alarm Battery Low Indicator. This indicator lights when the 

charge on the alarm battery is low. Press this button to 

display the percent of full battery charge in the digital display 
window.

19. Voltage Boost Indicator. This indicator lights when incoming 

voltage is low and the Voltage Safeguard has been activated 
(refer to Section 7 on page 7). Press this pad to display the 
current line voltage in the digital display window.

23. Alarm On/Set Point Security Indicator. This indicator lights 

when the alarm is activated.Also indicates that Setpoint 
Security is activated.

The following indicators are available only when the freezer is 
equipped with the optional backup system (refer to Section 6.3.4 
on page 6).

20. Backup System Supply On Indicator.
21. Backup System Empty Indicator
22. Backup System Battery Low Indicator.

off

life-guard alert

extreme ambient alert

clean filter

power failure

temperature failure

alarm testalarm reset

cabinet

temperature

control

set point 

alarm

battery low

voltage 

warm alarm

set point

cold alarm

set point

power on

alarm on

set point security

Ultima

1

15

7

9

10

11

SUPPLY

ON

SUPPLY

EMPTY

BACKUP SYSTEM

BATTERY

LOW

8

23

6

12

13

21

20

22

14

18

19

17

16

2

5

4

3

check

 fuse

boost

 


background image

Thermo Scientific Revco Cryogenic Storage Freezers

Installation and Operation

4

6.2 Start Up

Refer to Section 6.1 and Figure 2 on page 3 as you complete the 
following procedures.

6.2.1 Turning the Power On

To start up your cryogenic freezer, complete the following steps:

1. Plug the freezer into the power outlet (refer to Section 4.2 on 

page 2).

2. Turn the key switch to the POWER ON position. The Power 

On LED and the Cabinet Temperature LED illuminate.

Note:

The alarm function is not active at this time. Refer to 
Section 5.2.3 for information about setting the alarms. In 

addition, the Setpoint Security will be deactivated in this 
key position.

3. The unit will start about 30 seconds later.

6.2.2 Setting the Cabinet Temperature

To set the cabinet temperature, complete the following steps:

1. Press and hold the Control Setpoint pad.

2. The Control setpoint LED lights and the Cabinet 

Temperature LED goes out.

3. Press and hold 

 to increase the temperature or 

 to 

decrease the temperature. The digital display scrolls through 
temperature settings.

4. Release both pads when the digital temperature display 

window shows the correct setpoint value.

Note:

If no keys are pressed within ten seconds, the 
temperature display reverts to the cabinet temperature.

6.2.3 Setting the Alarms

To set the cold alarm, complete the following steps:

1. Confirm that the key switch is at the Power On position 

(Setpoint Security disabled) 

2. Press and hold the Cold Alarm Setpoint pad. The LED next 

to this pad lights. The temperature display shows the Cold 

Alarm value.

3. Press and hold 

 or 

 to adjust the Cold Alarm Setpoint.

4. Release both pads when the digital temperature display 

window shows the correct setpoint value.

To set the warm alarm, complete the following steps:

1. Confirm that the key switch is at the Power On position 

(Setpoint Security disabled).

2. Press and hold the Warm Alarm Setpoint pad. The LED next 

to this pad lights. The temperature display shows the Warm 

Alarm value.

3. Press and hold 

 or 

 to adjust the Warm Alarm Setpoint.

4. Release both pads when the digital temperature display 

window shows the correct setpoint value. 

When the cabinet temperature drops below the Warm Alarm 
setting, turn the key switch to ALARM ON. The freezer is ready 
to operate.

Note:

If a power failure lasting over 30 seconds occurs, the 

POWER ON light goes out and the POWER FAILURE 
light and digital temperature display flash 
simultaneously.

WARNING!

You must turn the three-position key 

switch to the ALARM ON/SETPOINT SECURITY 
position to activate the alarm and place it in security 
operation.

6.2.4 Alarm Reset & Status Lights 

The alarm reset feature ensures user acknowledgment of the 
occurrence of certain alarm conditions. This provides greater 
monitoring power by alerting the user if an alarm condition has 

occurred during periods when the freezer must be left 
unattended. The alarm conditions incorporating this feature are:

Temperature Failure-Warm Alarm

Temperature Failure-Cold Alarm

Voltage Low

Power Failure

Life-Guard Alert

Extreme Ambient Alert 

During any of these alarm conditions, the corresponding 
indicator will flash quickly, approximately 90 times per minute. 

If the alarm condition then disappears, the flash rate will 
decrease to approximately 15 times per minute. The indicator 
will then remain flashing at this rate, unless the condition 

reoccurs, until the Alarm Reset button is pressed. Once this has 
occurred, the indicator will no longer be lit.

In the case of TEMPERATURE FAILURE-WARM ALARM and 

TEMPERATURE FAILURE-COLD ALARM, if these 
conditions have occurred but are now within the given limits, the 
two indicators will alternately flash (180° out of phase). In 

addition, for these two temperatures failures, the highest 
temperature and lowest temperature, respectively, during the 
error condition will be saved. Prior to the temperature failure 

being reset, the temperature extreme may be viewed by pressing 
the Cabinet Temperature pad in conjunction with either the 
Warm Alarm Setpoint pad or the Cold Alarm Setpoint pad.

Note:

Once the Alarm Reset button is depressed, the extreme 
temperature value is reset to the current cabinet 
temperature.

6.2.5 Alarm Test 

To test the cold and warm alarms, complete the following steps:

1. Press and hold the Alarm Test pad. The digital temperature 

display indicates rising temperature. When the temperature 
reaches the warm alarm value, the alarm sounds.

Note:

The temperature of the refrigerated space does not 
change during this procedure. Only the sensor is heated.

2. Release the Alarm Test pad, the sensor and display return to 

cabinet temperature in a few minutes.

 


background image

Thermo Scientific Revco Cryogenic Storage Freezers

Installation and Operation

5

You can press the Audio Silence pad to silence the alarm but the 

alarm will ring back in five to seven minutes. You can silence the 
alarm in this manner as many times as necessary until the sensor 
cools below the warm alarm temperature. The alarm quits 
automatically when the sensor cools below the warm alarm 

value.

Note:

There is a limit on the alarm test feature. The alarm test 
will not commence if the warm alarm setpoint is more 

than 15 degrees warmer than the operating temperature 
setpoint.

6.3 Backup System (Optional)

When you purchase a built-in LN

2

 optional backup system for 

the freezer, the backup system control panel is located behind a 
hinged panel adjacent to the grill (refer to Figure 3).

Note:

Always purchase the cylinders which are equipped with 

siphon tubes for withdrawing liquid from the bottom of 
the cylinder. LN

bottles are functional at any reasonable 

temperature.

Use this panel to set the backup initiation temperature, usually 
10 to 15

°

C warmer than the operating setpoint.

Figure 3.  Control Panel for Optional 

Built-In Backup System

6.3.1 LN

2

 Precautions

The following are precautions for using liquid LN

2

 backup 

systems.

WARNING!

If an LN

2

 cylinder falls and a valve is 

knocked off, the cylinder becomes a deadly and 
completely unguided missile. Transport the cylinders 
in a handtruck or cart with secure chain ties for the 
cylinder. After cylinders are connected to the 
equipment, securely attach them with chains to a 
solid, stationary object such as a building column.

WARNING!

LN

2

 liquid is non-poisonous but it is 

very cold and will burn unprotected skin. Always wear 
protective eyewear and clothing when changing 
cylinders or working on the piping systems attached to 
an active source of liquid refrigerant.

WARNING!

 The gases produced by evaporation of 

LN

2

 are non-poisonous but displace the oxygen in a 

confined space and can cause asphyxiation. 

Do not 

store the cylinders in subsurface or enclosed areas

.

CAUTION!

When closing the cylinder valve, make 

sure that the injection solenoid is energized to allow all 
the liquid to bleed off instead of being trapped in the 
supply tubing. Failure to do this results in activation of 
the pressure relief device, which could damage the 
freezer and requires replacing if it is activated.

CAUTION!

For chest models ordered with factory 

installed built-in backup systems, the flow of liquid LN

2

 

will be discontinued if the door or lid is opened during 
operation of the backup system. For units operated 
with free-standing, field installed type backup system, 
the flow of liquid LN

2

 will be discontinued upon door or 

lid opening 

only

 if the switch provided with the 

free-standing package is installed on the freezer.

TEMPERATURE °C

PUSH

TO SET

SETPOINT

ADJUST

POWER

OFF

ON

Power

Switch

Temperature Control

Setpoint Screw

Temperature Control

Set Button

Digital Temperature

Display

 


background image

Thermo Scientific Revco Cryogenic Storage Freezers

Installation and Operation

6

6.3.2 Installation

Field installed systems are supplied with complete installation 
and operating instructions. If your system is factory installed, the 
freezer is shipped with a coiled length of tubing to connect the 
freezer to the bottles:

5/8 in. OD copper tubing covered with Armaflex

 insulating 

tubing for connection to the LN

supply.

Straighten the coiled tubing and connect one end to the labeled 

connection on the freezer and the other end to the supply bottle or 
building supply fitting. 

6.3.3 Operation

To activate the optional backup system, complete the following 
steps:

1. Push the Power Switch to the ON position. The digital 

temperature display window shows the current cabinet 
temperature.

2. Press and hold the Push to Set button to display the setpoint 

temperature in the digital temperature display window.

3. Use a small screwdriver to rotate the Setpoint Adjust screw 

until the correct backup operating temperature shows in the 

display window. The recommended setpoint is 10 to 15

°

above normal cabinet temperature.

Note:

The temperature shown in the display window is 
accurate to 

±

1

°

C.

The backup system can run for a minimum of eight hours on 
battery power.

WARNING!

You must push the Power Switch to the 

ON position in order to activate the backup system.

6.3.4 System Status Lights

In addition to the backup control system, the built-in backup 
system includes main control panel system status lights. These 
lights are only available when the backup system is installed.

The Supply On Indicator lights when the backup system is 
operating on battery power.

The Backup System Battery Low Indicator lights when the 
backup system battery charge is low. The built-in battery 
charger recharges the battery to full charge when power is 

restored to the system.

Note:

Since rechargeable batteries degrade over time, the 
battery should be replaced after approximately three 

years.

The Backup System Empty Indicator lights when the standby 
LN

2

 bottle is empty.

Figure 4.  Location of LN

2

 Connection Site, 

Chest Models Only

Table 1. 

Backup System Flow Rates – Cryogenic Freezers

Lbs/hr @ 75

×

Ambient Temperature 

(Flow Rate)

Center Air 
Temperature

-150

×

C

-140

×

C

-130

×

C

-120

×

C

Liquid LN

2

Chest Freezers 

7 ft

3

Empty

11.3

10.0

9.0

8.0

1/2 Full

9.7

8.6

7.6

6.8

Full

7.8

6.8

6.0

5.2

10 ft

3

Empty

15.2

13.5

11.9

10.3

1/2 Full

12.0

10.4

9.4

8.3

Full

8.9

7.8

7.0

6.1

24.75"

2.375"

LN

2

 Inlet Conn.

Copper Tubing

Factory Supplied

0.625" O.D.

MACHINE COMPARTMENT

REAR VIEW

 


background image

Thermo Scientific Revco Cryogenic Storage Freezers

Installation and Operation

7

7 Maintenance and Troubleshooting

WARNING!

Unauthorized repair of your freezer will 

invalidate your warranty. Contact Technical Service at 
1-800-438-4851 for additional information.

CAUTION!

Maintenance should only be performed 

by trained personnel.

7.1

Condenser Maintenance

7.1.1 Cleaning the Condenser

Clean the condenser at least every six months; more often if the 

laboratory area is extremely dust prone. 

To clean the condenser, complete the following steps:

1. Pull the grill open.

2. Remove the filter. Check the fans. If a fan is not operating, 

contact an Authorized Service Company immediately.

3. Vacuum the condenser.

4. Replace the filter and close the grill.

7.1.2 Cleaning the Condenser Filter

Clean the condenser filter every two or three months.

1. Pull the grill open.
2. Remove the filter.
3. Shake the filter to remove loose dust, rinse the filter in clean 

water, shake the excess water from the filter, and replace the 
filter.

4. Close the grill.

7.2

Gasket Maintenance

Periodically check the gaskets around the door or lid for 
punctures or tears. Leaks are indicated by a streak of frost which 

forms at the point of gasket failure. Make sure that the cabinet is 
level (refer to Section 5.3 on page 2 for leveling information).

Keep the lid gasket clean and frost free by wiping gently with a 

soft cloth.

7.3

Defrosting the Freezer

Defrost the freezer once or year or whenever the ice buildup 
exceeds 3/8”.

To defrost, complete the following steps:

1. Remove all products and place in another cabinet.
2. Turn off the freezer.

3. Open the outer door and all inner doors.
4. Let the freezer stand with the door open for at least 24 hours. 

This allows both the interior and foamed refrigerant system 
to warm to room temperature.

5. Dispose of the ice and wipe out any water standing in the 

bottom of the cabinet.

6. If there is freezer odor, wash the interior with a solution of 

baking soda and warm water. Clean the exterior with any 
common household cleaning wax.

7. Close the doors, restart the freezer and reload, following the 

instructions in Section 3.1 on page 1.

7.4

Alarm Battery Maintenance

Have a technician check the condition of the alarm battery at 
least once a year.

To replace the alarm battery, complete the following steps:

1. Remove the front grill. The alarm battery is located directly 

behind the grill. The terminals are the “push on” type.

2. Grasp the terminal with pliers and work it gently back and 

forth while pulling it off. The fittings are tight.

3. Remove the battery and put the new battery in place.

Note:

You may have to cut a strip of silicone rubber in order to 
remove the battery.

4. Connect the battery terminals and replace the front grill.

7.5

Frame Heater Switch

Cryogenic chest freezers (-140º and -150º C) have a frame heater 
switch which can be used to prevent or remove condensation 
near the freezer compartment lid. The switch is located in back of 

the unit near the top center, to the right of the junction box. 
Turning the switch to the up position heats up the frame around 
the lid. 

Heating the frame has no effect on interior cabinet temperature. 
You can turn on the switch whenever there is condensation near 
the lid.

7.6

Troubleshooting Procedures

Table 2. 

Troubleshooting Procedures

Problem

Solution

Check Fuse 
indicator is on.

Surge protector is blown. Call an Electronics 
Technician to check the equipment.

Clean Filter 
indicator is on.

Condenser filter is dirty. Refer to Section 7.1.2.

Alarm Battery Low 
indicator is on.

Alarm battery is low on charge. Have a 
technician check the alarm battery. Refer to 
Section 7.4.

Voltage Boost 
indicator is on.

Low incoming line voltage to the freezer. Call an 
electronics technician to check the power supply.

No display, no fans 
running.

No power to the unit. Call a qualified electrical 
service technician to check the service breaker 
to the outlet.

 


background image

Thermo Scientific Revco Cryogenic Storage Freezers

Installation and Operation

8

8 Chart Recorders 

Panel-mounted six-inch seven-day recorders are available as 
options for all freezer models.

8.1 Set Up and Operation

To prepare the recorder to function properly, complete the 

following steps:

1. Open the recorder door to access the recorder.
2. Connect the nine volt DC battery located at the recorder’s 

upper right corner. This battery provides backup power.

3. Install clean chart paper (refer to Section 8.3 below).
4. Remove the plastic cap from the pen stylus and close the 

recorder door.

Recorder operation begins when the system is powered on.The 
recorder may not respond until the system reaches temperatures 
within the recorder’s range. 

Figure 5.  Chart Recorder

Figure 6.  Chart Buttons

8.2 Power Supply

The recorder normally uses AC power when the system is 
operating. If AC power fails, the LED indicator flashes to alert 
you to a power failure. The recorder continues sensing cabinet 
temperature and the chart continues turning for approximately 

24 hours with back-up power provided by the nine-volt battery. 

The LED indicator glows continuously when main power is 
functioning and the battery is charged.

When the battery is low, the LED flashes to indicate that the 
battery needs to be changed.

8.3 Changing Chart Paper

To change the chart paper, complete the following steps:

1. Locate the pressure sensitive buttons at the front, upper left 

of the recorder panel.

2. Press and hold the Change Chart button (#3) for one second. 

The pen will move off the scale.

3. Unscrew the center nut, remove the old chart paper, and 

install new chart paper. Carefully align the day and time with 

the reference mark (a small groove on the left side of the 
recorder panel).

4. Replace the center nut and hand tighten. Press the Change 

Chart button again to resume temperature recording.

8.4 Calibration Adjustment

This recorder has been accurately calibrated at the factory and 
retains calibration even during power interruptions. If required, 

however, adjustments can be made as follows:

1. Run the unit continuously at the control setpoint temperature. 

Continue steady operation for at least two hours to provide 

adequate time for recorder response.

2. Measure cabinet center temperature with a calibrated 

temperature monitor.

3. Compare the recorder temperature to the measured cabinet 

temperature. If necessary, adjust recorder by pressing the left 
(#1) and right (#2) chart buttons.

Note:

The stylus does not begin to move until the button is held 

for five seconds.

If you find that the recorder frequently needs to be calibrated 
after paper changes, the program may have changed. To confirm 

that the program is correct:

1. Press button #3 and allow the pin to move off the scale.
2. Press #1 for 10 seconds and release.

3. Confirm that the green LED flashes 3 times followed by a 

pause. If the number of blinks is greater than or less than 3, 
adjust the number using the left (#1) and right (#2) chart 

buttons.

4. When the green LED blinks 3 times to confirm the resetting, 

press the #3 button. The pin arm should then return to 
indicate the current temperature. Repeat Steps 1-3 if 

necessary.

9 Volt Battery

I

mprinting

Stylus

Hub-Nut and
Retaining Wire

Chart

Chart Buttons

Reference 
Mark

1

2

3

CHANGE CHART

 


background image

9

Thermo Scientific Revco Cryogenic Storage Freezers

Installation and Operation

WEEE Compliance

WEEE Compliance

. This product is required to comply with the European 

Union’s Waste Electrical & Electronic Equipment (WEEE) Directive 2002/96EC. 
It is marked with the following symbol. Thermo Fisher Scientific has contracted 
with one or more recycling/disposal companies in each EU Member State, and this 
product should be disposed of or recycled through them. Further information on 
Thermo Fisher Scientific compliance with these Directives, the recyclers in your 
country, and information on Thermo Scientific products which may assist the 
detection of substances subject to the RoHS Directive are available at 

www.thermo.com/

WEEE Konformittät

. Dieses Produkt muss die EU Waste Electrical & Electronic 

Equipment (WEEE) Richtlinie 2002/96EC erfüllen. Das Produkt ist durch 
folgendes Symbol gekennzeichnet. Thermo Fisher Scientific hat Vereinbarungen 
getroffen mit Verwertungs/Entsorgungsanlagen in allen EU-Mitgliederstaaten und 
dieses Produkt muss durch diese Firmen widerverwetet oder entsorgt werden. 
Mehr Informationen über die Einhaltung dieser Anweisungen durch Thermo 
Fisher Scientific, die Verwerter und Hinweise die Ihnen nützlich sein können, die 
Thermo Scientific Produkte zu identifizieren, die unter diese RoHS. 
Anweisungfallen, finden Sie unter 

www.thermo.com/

Conformità WEEE. 

Questo prodotto deve rispondere alla direttiva dell’ Unione 

Europea 2002/96EC in merito ai Rifiuti degli Apparecchi Elettrici ed Elettronici 
(WEEE). È marcato col seguente simbolo.Thermo Fisher Scientific ha stipulato 
contratti con una o diverse società di riciclaggio/smaltimento in ognuno degli Stati 
Membri Europei. Questo prodotto verrà smaltito o riciclato tramite queste 
medesime. Ulteriori informazioni sulla conformità di Thermo Fisher Scientific con 
queste Direttive, l’elenco delle ditte di riciclaggio nel Vostro paese e informazioni 
sui prodotti Thermo Scientific che possono essere utili alla rilevazione di sostanze 
soggette alla Direttiva RoHS sono disponibili sul sito 

www.thermo.com

/

Conformité WEEE. 

Ce produit doit être conforme à la directive européenne 

(2002/96EC) des Déchets d’Equipements Electriques et Electroniques (DEEE). Il 
est marqué par le symbole suivant. Thermo Fisher Scientific s’est associé avec une 
ou plusieurs compagnies de recyclage dans chaque état membre de l’union 
européenne et ce produit devraitêtre collecté ou recyclé par celles-ci. Davantage 
d’informations sur laconformité de Thermo Fisher Scientific à ces directives, les 
recycleurs dans votre pays et les informations sur les produits Thermo Scientific 
qui peuvent aider le détection des substances sujettes à la directive RoHS sont 
disponibles sur 

www.thermo.com/

Deutschland

Italia

France

Great Britain

 


background image

Important

For your future reference and when contacting the factory, please have the 

following information readily available:

Model Number:

Serial Number: 

Date Purchased: 

The above information can be found on the dataplate attached to the 

equipment. If available, please provide the date purchased, the source of 

purchase (manufacturer or specific agent/rep organization), and purchase 

order number.

IF YOU NEED ASSISTANCE:

SALES DIVISION

Phone: