PSR-1100 

Portable Spectroradiometer 

Operator’s Manual 

 
 

 

 

 


background image

 

©2012 Spectral Evolution Inc. 90 Sutton St. Unit 4, North Andover, MA 01845 USA 

PSR-1100 Manual 

Revision: 1.00 

4/5/2012 

 

Table of Contents 

1

 

Introduction............................................................................................................. 3

 

2

 

Hardware and Interface Description........................................................................ 4

 

2.1

 

Part Identification ............................................................................................ 4

 

2.1.1

 

Spectrometer........................................................................................... 4

 

2.1.2

 

Battery Pack ............................................................................................ 4

 

2.1.3

 

Optical Input ............................................................................................ 4

 

2.1.4

 

Serial Communications............................................................................ 4

 

2.1.5

 

Front Panel.............................................................................................. 5

 

2.1.6

 

Back Panel .............................................................................................. 5

 

2.1.7

 

Laser Scan Switch................................................................................... 5

 

2.2

 

Interface Specification ..................................................................................... 6

 

2.2.1

 

Battery Pack ............................................................................................ 6

 

2.2.2

 

USB Serial Connector ............................................................................. 6

 

2.2.3

 

Serial Communications via USB or Bluetooth Virtual COM Port .............. 6

 

2.2.4

 

ESD protection: ....................................................................................... 7

 

3

 

Control Panel .......................................................................................................... 8

 

3.1

 

Indicators ........................................................................................................ 8

 

3.2

 

Switches ......................................................................................................... 8

 

3.2.1

 

Power Switch........................................................................................... 8

 

3.2.2

 

Membrane Switches ................................................................................ 8

 

3.2.3

 

Switch Function ....................................................................................... 8

 

3.3

 

LCD Display .................................................................................................... 8

 

3.3.1

 

Scan Types ............................................................................................. 8

 

3.3.2

 

Status Messages..................................................................................... 9

 

3.3.3

 

Parameter Menu...................................................................................... 9

 

3.3.4

 

Modifying a Parameter .......................................................................... 10

 

4

 

Operational/Functional Description ....................................................................... 11

 

4.1

 

Setup ............................................................................................................ 11

 

4.2

 

Standalone Operation ................................................................................... 11

 

4.3

 

Operation with a host PC .............................................................................. 13

 

5

 

Battery Charger .................................................................................................... 14

 

5.1

 

Introduction ................................................................................................... 14

 

5.2

 

Operation ...................................................................................................... 14

 

5.3

 

Information and Warnings ............................................................................. 14

 

Appendix A: Laser Safety ............................................................................................. 15

 

Appendix B:  PSR-1100 Specifications ......................................................................... 17

 

Appendix C: Bluetooth Information ............................................................................... 18

 

 

 

 


background image

 

©2012 Spectral Evolution Inc. 90 Sutton St. Unit 4, North Andover, MA 01845 USA 

PSR-1100 Manual 

Revision: 1.00 

4/5/2012 

 

1  Introduction 

 

This manual contains data concerning the PSR-1100 field-portable 

spectroradiometer system.  The system consists of a UV-visible-NIR range spectrometer 
with integrated lens (or optional fiber optic input)  that produces output calibrated to units 
of spectral radiance (W/m

2

/nm/sr). Energy enters the spectrometer from an external 

illumination source by way of the lens.  The spectrometer is a dispersive PDA system 
operating in the range of 320–1100 nm with a 512-element array detector.   
 
 

All electronics for controlling the PSR-1100 system operation are built into the 

instrument. The instrument incorporates manual controls and on-board storage for self-
contained field operation, as well as the capability to connect to a host computer system 
using the provided application software via USB or Bluetooth communications interface. 
The instrument is powered by a Li-Ion battery pack (provided with each system). 
 
 

The PSR-1100 contains an internal microprocessor that controls the data 

acquisition of the detector array, the communication interface to the host PC, and 
provides data storage.   Calibration data and spectral data (collected in standalone mode, 
described below) are stored in flash memory. The PSR-1100 has a built-in shutter for 
automatic dark measurements. The dark current data are retained in RAM and the 
microprocessor automatically performs mathematical functions on collected spectral 
scan data, including dark subtraction and automatic integration adjustment.  
 

 

 

 

 


background image

 

©2012 Spectral Evolution Inc. 90 Sutton St. Unit 4, North Andover, MA 01845 USA 

PSR-1100 Manual 

Revision: 1.00 

4/5/2012 

 

2  Hardware and Interface Description 

2.1  Part Identification 

2.1.1  Spectrometer 

 

The spectrometer component of the SR-1100 is a crossed Czerny-Turner 

configuration using a ruled grating as the dispersive element.  Energy enters the 
spectrometer and is collimated before being reflected off the grating and refocused onto 
the PDA detector.  The detector is a 512-element array, covering the spectral range 
320–1100 nm.  The spectrometer and controlling electronics are contained in the 
housing. 

2.1.2  Battery Pack 

The PSR-1100 uses a Li-Ion battery pack as a power source, supplied with the 

system. The Li-Ion battery runs at a nominal 7.4V and is designed to provide more than 
3 hours operation following a full charge. When the battery voltage falls too low to power 
the instrument, it will turn off and go to 0 volts. Remove the battery from the instrument 
for recharging with the supplied charger. Refer to 

Section

 

0:  

 

 


background image

 

©2012 Spectral Evolution Inc. 90 Sutton St. Unit 4, North Andover, MA 01845 USA 

PSR-1100 Manual 

Revision: 1.00 

4/5/2012 

 

Battery

 Charger for charging details. 

 

2.1.3  Optical Input 

 

The PSR-1100 comes with a standard fixed 4° field-of-view lens foreoptic 

installed. Custom configurations may replace the lens with alternate foreoptics such as a 
25° FOV fiber optic input. 
 
 

The entire spectrometer/foreoptic system is factory-calibrated for radiance using 

a NIST-traceable source. The calibration coefficients are stored in flash memory by the 
internal microprocessor. The application software retrieves these coefficients and 
applies them to acquired scan sample data to convert to radiometric units. 

2.1.4  Serial Communications 

 

Communications with the PSR-1100 are conducted using USB or Bluetooth.  

 
 

USB communications require a standard cable with USB-B connector to fit the 

receptacle on the control panel. The first time the host PC is connected to the instrument, 
Windows® will need to install drivers for the virtual COM port functionality. These drivers 
are included on the CD-ROM that was provided with your unit. Once the drivers are 
loaded, the provided Windows® application software interacts with the 
spectroradiometer using a virtual COM port in the same fashion as a standard RS-232 
serial port (115200 baud, 8 data bits, 1 stop bit, no parity).   
 

The PSR-1100 is also enabled for wireless communications with the application 

software provided the user has installed a Bluetooth serial port on the host PC. 
Depending on your Bluetooth adapter’s configuration utility, you will need to enter an 
identifier for your instrument type (typically called its PIN key or pairing code) the first 
time you use your PSR-1100. This code is “psr1100”. Your Bluetooth utility will recognize 
the instrument by a unique name with the syntax “PSR-1100 ####”, where #### is 
replaced by the serial number of the unit. 

 

 


background image

 

©2012 Spectral Evolution Inc. 90 Sutton St. Unit 4, North Andover, MA 01845 USA 

PSR-1100 Manual 

Revision: 1.00 

4/5/2012 

 

 

Figure 1: Front Panel view of the PSR-1100 

2.1.5  Front Panel 

The front panel of the PSR-1100 contains the on/off switch, USB connector, 

battery receptacle and a control panel with membrane switches and LCD display. A 
handstrap is permanently attached and near the front panel are two mounting brackets 
for attaching a shoulder strap. See 

Section 3: Control Panel 

for more information. 

2.1.6  Back Panel 

 

The opposite end of the PSR-1100 contains the optical input (lens or SMA fiber 

optic input), Bluetooth antenna and optional tripod mount. 

2.1.7  Laser Scan Switch 

 

The Laser Scan switch is located on the top cover of the spectroradiometer. 

Depressing the switch actuates the sighting laser and can be optionally configured to 
initiate the start of a standalone spectral scan. 
 

DANGER: DO NOT LOOK INTO THE LASER BEAM AT ANY TIME, INCLUDING 
INSTRUMENT SETUP OR OPERATION.  
 

For more information see 

Appendix A: Laser Safety

 

Li-Ion 
Battery 

USB 
Connector 

On/Off Switch 

Control Panel 

Laser Scan Switch 

 


background image

 

©2012 Spectral Evolution Inc. 90 Sutton St. Unit 4, North Andover, MA 01845 USA 

PSR-1100 Manual 

Revision: 1.00 

4/5/2012 

 

 

Figure 2: View of PSR-1100 optical components 

2.2  Interface Specification 

2.2.1  Battery Pack 

 

Your PSR-1100 is supplied with one or more Li-Ion battery packs that fit into the 

recess on the front panel. Slide the spring-loaded catch to the side (away from the 
recess) with a finger to insert or remove the battery. 

 

Important

: See 

Section 0:  

 

Laser 

Lens 

Bluetooth 
Antenna 

 


background image

 

©2012 Spectral Evolution Inc. 90 Sutton St. Unit 4, North Andover, MA 01845 USA 

PSR-1100 Manual 

Revision: 1.00 

4/5/2012 

 

Battery

 Charger

 

for details on charging the external battery and safe operation. 

2.2.2  USB Serial Connector 

 

For communications via USB, the front panel is equipped with a USB Type-B 

receptacle connector, compatible with all standard USB cables with Type-B plug.  

Pin 1:  +5V 

 

Pin 2:  D- 

 

Pin 3:  D+ 

 

Pin 4:  GND 

2.2.3  Serial Communications via USB or Bluetooth Virtual COM Port 

Baud Rate:  115200 bps 
Parity: none 
Data Length: 8 
Stop bit: 1 
Flow Control: none 

2.2.4  ESD protection:  

 

+/- 15 KEV human body model,  

 

+/- 8KEV contact discharge  

 


background image

 

©2012 Spectral Evolution Inc. 90 Sutton St. Unit 4, North Andover, MA 01845 USA 

PSR-1100 Manual 

Revision: 1.00 

4/5/2012 

 

 

3  Control Panel 

 

The control panel of each PSR-series spectroradiometer contains the controls 

and displays that are required for the stand-alone operation of the instrument. 

3.1  Indicators 

The Control Panel has 3 LED status indicators. 
 

Power LED:   

The power indicator is illuminated when the power is on. 

Laser LED: 

 

The laser indicator is illuminated when the laser is on. 

Connect LED: 

The connect indicator is illuminated when the instrument is NOT 

connected to a computer via Bluetooth communication link. 

3.2  Switches 

3.2.1  Power Switch 

The power switch provides on/off control of the unit. 

3.2.2  Membrane Switches 

The three (3) membrane push button switches provide control of the unit during 
standalone operation and the capability of modifying the unit’s operation. The functions 
of the switches are detailed below. 

3.2.3  Switch Function 

Scan: 

Steps through scan options (Reference/Target/Dark Scans) 

Menu: 

Steps through the menu options (Memory Scan#/Dark Scan Mode/Scan 

Timer/Optic/Laser Scanning) 

Edit: 

Toggles or cycles through parameter lists for each menu option 

3.3  LCD Display 

The LCD display is an eight (8) character, two (2) row display that shows the selected 
scan and status messages. 
 

ROW 1 

displays the scan types, status messages, and battery voltage. The SCAN 

button allows the user to select the type of scan to perform. 

ROW 2 

displays the parameter menu. The MENU button allows the user to scroll 

through a set of available parameters, and the EDIT button changes the parameter 
settings. 
 

3.3.1  Scan Types 

The next type of scan to be collected is displayed in row 1 as follows: 
 

 


background image

 

©2012 Spectral Evolution Inc. 90 Sutton St. Unit 4, North Andover, MA 01845 USA 

10 

PSR-1100 Manual 

Revision: 1.00 

4/5/2012 

 

REF x.xx - 

Indicates that a spectral measurement will be stored as a Reference Scan; 

current battery voltage is x.xx volts, or “LBAT” (Low BATtery) when indicated.  
 

TAR x.xx - 

Indicates that a spectral measurement will be stored as a Target Scan; 

current battery voltage is x.xx volts, or “LBAT” (Low BATtery) when indicated. 
 

DRK x.xx - 

Indicates that a dark measurement will be acquired. The dark measurement 

will be stored in volatile memory (not flash) and used for subsequent scaled dark 
measurements. The current battery voltage is x.xx  volts, or “LBAT” (Low BATtery) when 
indicated. 
 
Pressing the SCAN button cycles through these options. Note that after a REF scan is 
taken, the instrument automatically switches to collecting TAR scans. 

3.3.2  Status Messages 

These messages are also displayed on row 1: 

 
SCANNING - 

Indicates that a spectral data acquisition is being performed. 

BUSY

 - Indicates that the instrument is performing a lengthy internal operation. 

3.3.3  Parameter Menu 

The parameter menu is displayed on row 2. The MENU button allows the user to cycle 
through the following parameters. The EDIT button changes the parameter values. 

  
MEM# xxx

 - Displays the memory slot that will hold the next acquired scan. The 

allowable values are from 1 to 500. The memory slot cannot be adjusted by the keypad, 
with the exception of completely erasing all memory.  
To erase all memory and set MEM# to one (1), the <MENU> and <EDIT> keys must be 
pressed simultaneously. 

IMPORTANT!

 This operation permanently erases all internally 

stored scans taken in standalone operation mode. 
 

DRK xxxx

 - Selects the dark subtraction mode: 

AUTO

 - Automatic Dark mode is selected. Dark current measurement will be acquired 

automatically prior to each scan. The dark currents will be acquired at the same settings 
as the target scan.  

SCLD

 - Scaled Dark mode is selected.  Dark Current for each scan will be scaled, based 

on the integration time and the Dark Scan that was acquired manually using the Dark 
scan. 

 
STIM xxx - 

STIM xxx is the scan averaging time. The allowable values are from 1 to 30. 

The length of time for a scan is xxx seconds, for each light and dark scan.  
 

O:xxxxxx - 

Entrance optic option. 

This must be set to the entrance optic in use for 

correct calibration factors to be applied

. There can be up to eight (8) optic calibration 

options stored in a single instrument. Some possibilities are listed below: 

(1) 

LENS 4 – 4 degree FOV lens 

(2) 

FIBER1 – standard 25 degree FOV fiber optic 

 


background image

 

©2012 Spectral Evolution Inc. 90 Sutton St. Unit 4, North Andover, MA 01845 USA 

11 

PSR-1100 Manual 

Revision: 1.00 

4/5/2012 

 

(3) 

LENS14 – 14 degree FOV lens 

(4) 

RAW DN – no calibration applied 

 

LSCN xxx - 

Laser Scan Switch function. This can be set to ON or OFF. When the Laser 

Switch is depressed, the sighting laser is (always) turned on. When LSCN is set to ON, a 
stand alone scan will be acquired when the switch is released. When LSCN is set to 
OFF, no scan will be acquired. 

3.3.4  Modifying a Parameter 

(1) 

Depress the MENU button to locate the parameter you wish to edit. 

(2) 

Modify the parameter by depressing the EDIT button. The push button switches are 

programmed to scroll the menu of available options. 

(3) 

Depress the MENU button again to save the selected parameter setting. 

 

 

 


background image

 

©2012 Spectral Evolution Inc. 90 Sutton St. Unit 4, North Andover, MA 01845 USA 

12 

PSR-1100 Manual 

Revision: 1.00 

4/5/2012 

 

4  Operational/Functional Description 

All Spectral Evolution PSR series spectroradiometers operate as portable field 

instruments with onboard data storage and an easy-to-use manual control panel for 
standalone operation. Combined with a host PC, the PSR can also be controlled for 
acquisition in real time and for retrieval and display of the stored scan data. 

4.1  Setup 

To use the PSR-1100, install a fully charged battery pack by inserting it into the battery 
slot located on the instrument’s front panel. Switch the power switch to the ON position. 
After a brief initial power-on period, the LCD indicator will display the message “REF” 
followed by a battery voltage reading and “O: LENS4”.

 

4.2  Standalone Operation 

Dark Measurement Overview 

A dark measurement must be taken prior to taking reference or target 

measurements. Dark measurements are taken automatically by the instrument in 
AutoDark Mode, or are taken manually by the operator when in Scaled Dark Modes. 

In AutoDark Mode, the internal shutter automatically closes and dark 

measurement taken immediately prior to the reference or target scan; there is no need to 
ever take a dark measurement manually. 

If Scaled Dark Mode is used, however, a dark measurement must be taken 

manually. A new dark measurement should taken periodically, more often under 
changing environmental conditions.  

Using AutoDark Mode ensures that the latest dark values are subtracted on 

every scan; however, enabling AutoDark doubles the amount of time the scan requires 
to complete. The dark measurement is subtracted from the Reference and Target 
measurements. Dark measurements are stored in the PSR’s RAM and are not saved 
when the power is cycled. 
 

Reflectance Measurement Overview 

A reflectance measurement takes the scan of a reference (white plate) and 

calculates the ratio of the target scan to that reference. Unless the instrument is set to 
AutoDark Mode (where it takes a new dark reading with every scan) - if any of the below 
conditions exist, a new dark measurement should be taken: 
 

 After instrument power-up and warm-up, but before acquiring any reference/target 

scans. 

 Every 3-5 minutes thereafter; more often if time and conditions allow, or if lighting 

or environmental conditions are changing. 
 

If any of the below conditions exist, a new reference measurement should be 

taken before a series of target measurements: 

 Beginning of a new set of scans. 

 Any change in the collection optics. 

 The instrument has been idle for an extended period, or has been turned off. 

 


background image

 

©2012 Spectral Evolution Inc. 90 Sutton St. Unit 4, North Andover, MA 01845 USA 

13 

PSR-1100 Manual 

Revision: 1.00 

4/5/2012 

 

 Lighting conditions have changed or are variable. 

 

When in doubt, take a reference measurement! 
 

Taking a DARK Measurement (Skip these steps in AutoDark Mode) 

1. Check that the LED Power Indicator is illuminated, indicating that the power is 
enabled. 
2. Press the Menu membrane switch unit the LSCN menu item appears; press the Edit 
membrane switch so that ON appears. 
3. Press the Menu membrane switch until the DRK menu item appears; press the Edit 
membrane switch so that SCLD appears. This enables the Scaled Dark Mode for 
standalone operation. 
4. Press the Scan membrane switch until the DRK x.xx scan message appears on the 
LCD. (x.xx indicates battery voltage) Check to see if battery is adequately charged. 
5. Press and release the Laser Scan button on the side of the unit. The instrument will 
close the shutter and take dark data for approximately 15 seconds. Since the shutter 
closes in order to take the dark measurement, no particular target needs to be sighted 
by the instrument optics during a dark measurement. Dark measurements do not take up 
a memory location within the instrument’s stand-alone scan memory. 
6. The dark measurement just taken is associated with all succeeding measurements 
until a new Dark measurement is established. Dark measurements are stored in volatile 
memory (not flash) and must be retaken whenever the instrument power is cycled. 

 

 
Taking a REFERENCE Measurement 

1. Check that the LED Power Indicator is illuminated, indicating that the power is 
enabled. 
2. Press the Menu membrane switch unit the LSCN menu item appears; press the Edit 
membrane switch so that ON appears. 
3. Press the Scan membrane switch until the REF x.xx scan message on the LCD. (x.xx 
indicates battery voltage) Check to see if battery is adequately charged. 
4. Place a reference at the point of measurement making sure to completely fill the field 
of view of the instrument. This reference measurement is associated with all succeeding 
measurements until a new reference measurement is established. 
5. Press and hold the Laser Scan button to activate the sighting laser and verify the 
placement of the reference plate relative to the laser spot as required to fill the field of 
view. 
6. Release the Laser Scan button to record the measurement. The LCD counter 
increments to display the next available memory location. 

 
Taking a TARGET Measurement 

1. Check that the LED Power Indicator is illuminated indicating that the power is enabled. 
2. Press the Menu membrane switch until the LSCN menu item appears; press the Edit 
membrane switch so that ON appears. 
3. Press the Scan membrane switch until the TAR x.xx scan message on the LCD. (x.xx 
indicates battery voltage) Check to see if battery is properly charged. 
4. Place a reference at the point of measurement making sure to completely fill the field 
of view of the instrument. This reference measurement is associated with all succeeding 
measurements until a new reference measurement is established. 

 


background image

 

©2012 Spectral Evolution Inc. 90 Sutton St. Unit 4, North Andover, MA 01845 USA 

14 

PSR-1100 Manual 

Revision: 1.00 

4/5/2012 

 

5. Press and hold the Laser Scan button to activate the sighting laser and verify the 
placement of the target to be measured relative to the laser spot as required to fill the 
field of view. 
6. Release the Laser Scan button to record the measurement. The LCD counter 
increments to display the next available memory location. 

Additional target 

measurements may be made by pressing and releasing the Laser Scan switch.

 

 

4.3  Operation with a host PC 

 

The PSR-1100 interacts with Spectral Evolution’s DARWin SP software 

(provided with your instrument) running on the host PC via USB or Bluetooth virtual 
COM port connection.  
 

The primary tasks that can be performed by running DARWin SP while 

connected to your spectroradiometer are: 

1.  Controlling the instrument remotely and collecting and displaying scan data in 

real time, with expanded control options, and 

2.  Retrieving and displaying scan data collected and stored during standalone 

operation. 

 

Be sure to install the software (and any USB/Bluetooth communications drivers 

as needed) prior to the first time you connect to the instrument.  
 

When running DARWin SP make sure your instrument has a charged battery and 

that it is powered and that the proper communications interface has been enabled. Once 
communications have been established, a brief initialization sequence follows and then 
your system is ready for use. 
 

For detailed information on how to use DARWin SP with your PSR Series 

spectroradiometer, see the DARWin SP User Manual included with your instrument. 

 

 


background image

 

©2012 Spectral Evolution Inc. 90 Sutton St. Unit 4, North Andover, MA 01845 USA 

15 

PSR-1100 Manual 

Revision: 1.00 

4/5/2012 

 

5  Battery Charger 

5.1  Introduction 

The provided battery charger is used to charge one lithium-ion battery (nominal 

7.5VDC, 2800 mA-h). The charger operates from 110/240 volt AC line supply. Follow all 
directions for proper, safe operations.  Only charge batteries indoors in a cool place. 

5.2  Operation 

To charge the external battery pack: 

1.  Remove the battery pack from the instrument. 
2.  Plug the charger’s round male plug into the battery pack’s female jack. 
3.  Plug the charger into an AC outlet. 

The charger’s 

RED

 LED will be illuminated during charging; when charging is complete 

the charger’s 

GREEN

 LED will be illuminated. 

 
Charging a battery from empty to full capacity (approximately 8.4VDC) will take 
approximately 2-3 hours. 

5.3  Information and Warnings 

Cells: 18650 Li-ION cells with IC protection are used inside the battery pack 
 
Protection: 4A max polyswitch is installed in the battery pack for double over-discharging 
protection. 
 
The battery will be automatically cut off to zero (0) volts when its voltage falls too low for 
PSR operation. 
 

WARNINGS: 

  Avoid heavy impact to the battery pack; this may cause the battery pack to 

malfunction. 

  Do not expose the battery pack to fire or extreme heat; this may cause the 

battery pack to malfunction. 

  Do not use any DC power supply or charger other than the dedicated battery 

charger supplied with your instrument; this may cause the battery pack to 
malfunction. 

 
Also, always power the PSR Series instrument with its battery only. 

NEVER

 use the 

dedicated battery charger to provide power to the instrument. 
 

 


background image

 

©2012 Spectral Evolution Inc. 90 Sutton St. Unit 4, North Andover, MA 01845 USA 

16 

PSR-1100 Manual 

Revision: 1.00 

4/5/2012 

 

Appendix A: Laser Safety 

The PSR Series Spectroradiometers use a Class 3R laser beam that complies with IEC 
60825-1, Ed. 1.2. The source of laser energy in the PSR is a nominal 3 mw, continuous, 
635 nm diode laser. No maintenance or adjustments are required in order to keep this 
product in compliance with the above specifications. 

 

DANGER:  

Do not look into the laser beam at any time including instrument setup or 
operation. 

 

The laser is operated by pressing the Laser Scan Switch located on the top surface of 
the instrument. See 

Section 2: Hardware and Interface Description

 for the location of 

this control. 

 

CAUTION - Use of controls or adjustments or performance of procedures other 
than those specified herein may result in hazardous radiation exposure. 
 

There are no user serviceable components within this device. The unit should be 
returned to the manufacturer in case of malfunction. 
 

LABELING: 

The following labels are required for the PSR Series instrument. Each label is described 
below. If the instrument does not have the appropriate labels, contact the manufacturer 
to ensure that your unit is in compliance. 
 

(1)  Explanatory Label  

 

 

 

 

 

 

 

Location – Bottom of instrument, on the product information label. 

 

(2)  Supplementary Explanatory Label 

 

 

 

Laser Power: 3 mW, Laser Wavelength 635 nm. 

 

 

Laser conforms to IEC 60825-1, Ed 1.2. 

 
 

Location – Bottom of instrument, on the product information label. 

 

(3)  Warning Label 

 

 

 

 

 

 

Location – Bottom of instrument, on the product information label 

 
 

 


background image

 

©2012 Spectral Evolution Inc. 90 Sutton St. Unit 4, North Andover, MA 01845 USA 

17 

PSR-1100 Manual 

Revision: 1.00 

4/5/2012 

 

(4)  Aperture Label 

 

 

 

 

 

 

 

Location - Next to the laser aperture. 

 

 
 
 
 
 
 
 
 
 
 
 
 
 

 


background image

 

©2012 Spectral Evolution Inc. 90 Sutton St. Unit 4, North Andover, MA 01845 USA 

18 

PSR-1100 Manual 

Revision: 1.00 

4/5/2012 

 

Appendix B: PSR-1100 Specifications  

 

Performance Specifications 

 

Spectral Range 

320–1100 nm 

Spectral Resolution 
nominal 

<= 3.2 nm 

Spectral Sampling Bandwidth 
nominal 

1.5 nm  

Detector 

512 element Si PDA  

A/D Converter 

16 bit 

Wavelength Accuracy 

.5 nm 

Wavelength Reproducibility 

.1 nm 

Integration Time 

8-2000 ms 

 

Physical Specifications

 

 

 

Power 

Supplied 2800 mA-h Li-Ion battery pack with 

charger, 7.5V nominal 

Size 

7” x 3.25” x 5.75” 

Weight 

< 4 lbs 

Communications Interface 

USB and Bluetooth virtual COM ports 

Foreoptics 

4° lens provided (other options available) 

Onboard Storage 

> 500 scans in stand-alone mode 

Targeting 

Internal red laser 

 

Environmental Specifications 

 
 

Operation 

 

Temperature 

15 - 35° C (note: wvl drift 18 - 27° C) 

Humidity 

15 – 90 % non - condensing 

Storage 

 

Temperature 

0 - 70° C 

Humidity 

15 – 90 % non - condensing 

 

 


background image

 

©2012 Spectral Evolution Inc. 90 Sutton St. Unit 4, North Andover, MA 01845 USA 

19 

PSR-1100 Manual 

Revision: 1.00 

4/5/2012 

 

 

Appendix C: Bluetooth Information 

Bluetooth Versions Supported: V1.1 
Bluetooth Class: Class I 
Bluetooth Connection Type: Virtual COM Port 
Bluetooth Connection Pin Key/Pairing Code: psr1100 
 
 
 