PDF User Manual

  1. Home
  2. Manuals
  3. Infa Warmer ATOM Manual

Infa Warmer ATOM Manual

Made by: Infa Warmer
Type: Manual
Category: Baby Accessories
Pages: 4
Size: 0.54 MB

 

Download PDF User Manual



Full Text Searchable PDF User Manual



background image

ATOM Infant Warmer 

 


background image

From the delivery room to NICU—Meets wide needs 
with substantial functions and features. 

Superior operation performance and functionality will support daily nursing care. 

Pre-heating function 

Pre-heating condition enables to treat a baby under warm 
condition right after delivery. 

CPR timer

 

Integrated CPR timer in compliance with the international 
resuscitation consensus. APGAR timer is also integrated.

 

Note: Information in this document is subject to change without notice. 

Far-infrared radiation 

High warming efficiency thanks to Far-infrared radiation.  

Adjustable LED light

 

Even if the canopy is tilted, baby’s skin color can be clearly 
checked thanks to the LED light.  Light intensity can be 
adjusted. 

 

Drawer 

Drawer is available as an option.  The drawer is closed very 
silently in order to minimize the sound stress to a baby. 

Baby Guard 

Thanks  to  rotary  damper,  the  baby  guard  can  be  opened 
very silently without touching.  

LED Light 

Canopy 

Add-on Unit 

As add-on unit for Infant Warmer, two kinds of resuscitation unit are available depending on user’s demand. 

Resuscitation Unit 104 

Resusci—flow  is integrated for safety 
resuscitation treatment. 
 

● 

Manometer 

● 

Resusci– flow 

● 

Suction unit 

● 

O2 Blender 

 

Resuscitation Unit 105 

Oxygen  Blender  and  manometer  is 
integrated.  
 

● 

Manometer 

● 

O2 Blender 

● 

Suction Unit 

X-ray tray 

Hi-Low Stand 

 


background image

Specifications 

Note: Information in this document is subject to change without notice. 

 Atom Infant Warmer Model 103    

 Power requirements  

 Customer specified  

   

 Power consumption : 700VA (Maximum)  

 Heater capacity  

 500W  

 Classification  

 Class I, Type BF  

 Dimensions  

 84(W) X 110(D) X 178-218(H)cm    (Mattress surface : 80-120cm)  

 Weight  

 105kg  

 Temperature control  

 Servo control (Manual control also available)  

 Skin temperature  

 Setting range : 34.0-38.0

  

   

 Display range : 30.0-42.0

  

 Heater output  

 Setting range : 0-100% (in 5% increments)  

   

 Indication range : 0-100 (in 10 levels)  

 Alarms  

 Skin temperature, skin temperature probe, check baby, sysytem failure, main failure, canopy state  

 Timer  

 Time display range : 0-60minutes 00 seconds (in 1 second increments)  

   

 APGAR mode /CPR mode switchable  

 Lighting  

 LED lamp     Adjustable illumination intensity 0-1200lux  

 Mattress platform angle  

 Inclination angle (back and forth) : +13°

-13°  

 Communication port  

 RS-232  

 Accessories  

 Skin temperature probe (yellow) ……………………………1  

   

 Dust cover ………………………………...………………….1  

   

 Mattress sheet ……………………………..…………………1  

 

 

 Pulse Oximeter Unit  

   

 Display range  

 1-100%SpO2  

 Averaging time  

 8, 4, 16 seconds (Selectable)  

 SpO2 alarm setting range  

 Upper limit : 50-99% (in 1% increments)  

   

 Lower limit : 45-95% (in 1% increments)  

 Pulse rate display range  

 25-240bpm  

 Pulse rate alarm setting range    Upper limit : 80-240bpm, OFF  

   

 Lower limit : 35-180bpm, OFF (in 5bpm increments)  

 Alarms  

 SpO2 high/low limit, pulse rate high/low limit, SpO2 sensor  

 

 

 Resuscitation unit 104  

   

  Manometer  

 Pressure indication : -2 

~ 

8 kPa (-20 

~ 

80cmH2O)  

 Suciton unit  

 Suction pressure control range : 0 

~ 

26.7 kPa (0 

~ 

200mmHg)  

 O2 Blender  

 Oxygen concentration control range : 21 

~ 

100%  

   

 Oxygen flow rate control range : 0 

~ 

15L/min.  

 Resusci-flow unit  

 PMAX setting range : 19 

~ 

58 kPa (20 

~ 

60cmH2O)  

   

 Default setting range : 39 kPa (40cmH2O)  

 

 

 Resuscitation unit 105  

   

  Manometer  

 Pressure indication : -2 

~ 

8 kPa (-20 

~ 

80cmH2O)  

 Suciton unit  

 Suction pressure control range : 0 

~ 

26.7 kPa (0 

~ 

200mmHg)  

 O2 Blender  

 Oxygen concentration control range : 21 

~ 

100%  

   

 Oxygen flow rate control range : 0 

~ 

15L/min.  

 


background image

 Power Requirement  

AC120V 

AC220-240V 

 SpO2 Masimo  

△ 

△ 

 SpO2 Nellcor  

△ 

△ 

  Weight Monitor  

△ 

△ 

 Optional Add-on Unit  

  Resuscitation unit 104 PV  

△ 

△ 

  Resuscitation unit 104 EV  

△ 

△ 

  Resuscitation unit 105 PV  

△ 

△ 

  Resuscitation unit 105 EV  

△ 

△ 

Manufacturer: 

Atom Medical Corporation

 

Exporter: 

Atom Medical International, Inc.

 

Iwakata Bldg., 3-18-16, Hongo, Bunkyo-ku, Tokyo, JAPAN 
Phone: + 81 3 3815 2941 
Fax: + 81 3 3812 9670/4080 
URL: http://www.atom-ami.co.jp 

 

Model Number  

Model 103 

<Note>  
 104 PV: T-Piece Resuscitation Unit with Blender and Suction Unit (connect to central piping system)  
 104 EV: T-Piece Resuscitation Unit with Blender and Suction Unit of ejector type to make suction operation possible where no suction line is available  
 105 PV: O2 Blender Unit and Suction Unit (connect to central piping system)  
 105 EV: O2 Blender Unit and Suction Unit of ejector type to makes suction operation possible where no suction line is available  

 : Add-on Unit